Introduction: This study aimed to develop a novel therapeutic strategy for treatment-resistant cancers based on nano-brachytherapy, using gold nanoparticles as carriers for the Auger-electron–emitting radionuclides 197mHg and 197Hg.
Methods: Five-nanometre PEGylated gold nanoparticles were functionalised with 197mHg/197Hg via surface amalgamation and evaluated in vitro in triple-negative breast cancer (MDA-MB-231) and glioblastoma (T98G) cell lines. Cellular internalisation and subcellular distribution were assessed by uptake studies and fractionation. Cytotoxicity was evaluated using MTS assays, while therapeutic efficacy in three-dimensional models was investigated in tumour spheroids. Flow cytometry was employed to analyse apoptosis and cell-cycle distribution. DNA double-strand breaks were quantified by γ-H2AX phosphorylation. Ex vivo biodistribution studies compared intratumoural and intravenous administration routes.
Results: The radio-conjugate underwent straightforward synthesis, exhibited very high radio-labelling efficiency at low mercury loading, and maintained excellent colloidal stability. Efficient cellular uptake and pronounced nuclear accumulation were observed in both cell lines. In vitro studies demonstrated strong, time- and dose-dependent cytotoxicity predominantly mediated by apoptosis, with minimal necrosis, accompanied by significant induction of DNA double-strand breaks. In three-dimensional cultures, MDA-MB-231 spheroids underwent rapid, dose-dependent disintegration, whereas T98G spheroids displayed increased resistance. Ex vivo biodistribution revealed high tumour retention following intratumoural administration, minimal systemic exposure and predominantly renal clearance.
Conclusions: 197mHg/197Hg-functionalised gold nanoparticles represent a promising receptor-independent platform for localised nano-brachytherapy and warrant further preclinical evaluation in aggressive and therapy-resistant tumours.
Objective: Neurotoxicity is undesirable property of substances, especially those reaching the central nervous system. In this study we investigated neurotoxicity by modelling two distinct endpoints: neuronal metabolic activity and electrical activity.
Methods: Based on data from in vitro experiments on human cell lines, we built predictive models using automated machine learning (AutoML) and deep symbolic optimisation (DSO), trained on six molecular representations.
Results: The mathematical equations generated using DSO outperformed AutoML models, offering superior predictive accuracy and an advantage in interpretability. The best results for electrical activity prediction were achieved with RDKit descriptors (R2 of 0.863 on the train set and 0.861 on the test set). For metabolic activity, the best model relied on MACCS fingerprints (R2 of 0.645 on the train set, 0.627 for the test set).
Conclusions: These findings highlight the potential of data-driven approaches to anticipate neurotoxicity and demonstrate that DSO can create accurate and transparent models to support the design of safer chemicals.
Objective: The goal of the work is to develop a set of methods for calibrating the positron emission tomography system built from plastic scintillators, and to present the results of the modular J-PET scanner time of flight (TOF) and time over threshold (TOT) synchronisation.
Methods: Measurements with radionuclide 22Na and 44Sc (used as a point-like source and enclosed in a collimator) were performed using the modular J-PET scanner, which detects signals from annihilation photons and prompt gamma. The data were gathered with a trigger-less data acquisition system and analysed with a dedicated software framework. The detection modules were synchronised with the procedures, correcting time offsets and normalising TOT measurements.
Results: The application of the time calibration methods yields a fully synchronised detector that is capable of producing the image of the source. TOF resolution for modular J-PET is determined to be about 490 ps (FWHM).
Conclusions: The J-PET scanner, built from plastic scintillators, can be calibrated using β+γ emitters and taking advantage of the fact that the direction of propagation of annihilation and prompt photons are not correlated.
Objective: Hypoxia is a critical factor in tumour aggressiveness, metastasis, and treatment resistance. Despite its clinical significance, a non-invasive, high-precision method for assessing and mapping tissue oxygenation in vivo remains a major challenge in medicine.This study explores the potential of positronium imaging and quantum entanglement (QE) imaging as next-generation biomarkers for hypoxia. This manuscript introduces a novel method to assess tissue oxygen concentration via the QE of photons originating from positronium – a bound state of an electron and a positron – which is produced within the patient’s body during positron emission tomography (PET). We also investigate the possibility of assessing hypoxia by simultaneously detecting positronium lifetime and the positronium decay rate ratio.
Methods: We introduce two distinct quantum sensing approaches. Method 1 utilises the correlation between oxygen concentration and ortho-positronium (o-Ps) decay rates, relying on the simultaneous measurement of the mean o-Ps lifetime (τoPs) and the 3γ-to-2γ annihilation rate ratio of o-Ps (RoPs-3γ/2γ). Method 2 introduces a novel hypothesis based on the degree of QE of annihilation photons. This method leverages recent discoveries indicating that the degree of QE is sensitive to the relative contribution of annihilation mechanisms (pick-off vs. conversion), which in turn depends on the oxygen concentration. We estimate the rate of conversion processes as a function of the partial pressure of oxygen (pO2) in water, isopropanol, cyclohexane, isooctane, and adipose tissue. We consider dependence of RoPs-3γ/2γ and τoPs, as well as the degree of QE, on oxygen pressure in the studied substances. Finally, we derive a formula for pO2 as a function of RoPs-3γ/2γ and τoPs and estimate the measurement accuracy required for these parameters – and for the degree of QE – to sense in-vivo oxygen pressure in the range between hypoxic and physoxic conditions.
Results: Theoretical models and quantitative estimates for RoPs-3γ/2γ, τoPs and for the degree of QE (CQE and RQE) as a function of pO2 are provided for water, organic solvents (isopropanol, cyclohexane, isooctane), and adipose tissue. Applying the formulas derived under the working hypothesis that in pick-off process the photons are not entangled, we estimated that for pO2 = 0, the degree of QE CQE is equal to 0.890 for adipose, 0.886 for isopropanol, 0.867 for water, 0.818 for cyclohexane, and 0.784 for isooctane. These results indicate that distinguishing between various tissue types requires a precision of σ(CQE) ≈ 0.01. We also estimated the values of τoPs, RoPs-3γ/2γ, and R3γ/2γ, as well as the changes in τoPs, RoPs-3γ/2γ and CQE between physoxic (pO2 ≈ 50 mmHg) and hypoxic (pO2 = 10 mmHg) conditions amount to ΔτoPs = 5 ps (water), 48 ps (adipose), 182 ps (isopropanol), 187 ps (cyclohexane), 444 ps (isooctane), ΔRoPs-3γ/2γ = 0.000045 (water), 0.0004 (adipose), 0.0015 (isopropanol), 0.0015 (cyclohexane), 0.0036 (isooctane), and ΔCQE = 0.0003 (water), 0.0019 (adipose), 0.0054 (isopropanol), 0.0106 (cyclohexane), 0.0243 (isooctane). To distinguish between physoxic and hypoxic conditions in vivo, the measurement precision σ(τoPs), σ(RoPs-3γ/2γ) and σ(CQE) must be several times higher than the predicted changes in these values.
Conclusions: This work establishes a theoretical and methodological foundation for using the QE of photons and the properties of positronium as transformative diagnostic tools for the non-invasive assessment and mapping of tissue oxygenation in the human body. It demonstrates that QE can serve not only as a tool for improving PET image quality through noise reduction, but also as a completely new category of biomarker. Quantitative estimations of the effect of oxygen pressure on positronium parameters, as well as the event statistics required for hypoxia assessment, are feasible using the multi-photon total-body J-PET scanner based on plastic scintillators. Furthermore, such assessments will be possible with next-generation, high-sensitivity crystal-based PET scanners, provided they are upgraded to support multi-photon signal acquisition for τoPs and 3γ-to-2γ imaging, and double Compton scattering for imaging CQE – the degree of QE.
Objective: This study aims to demonstrate that cellular heterogeneity, typically associated with cancer cells, also occurs in normal cells. The study also investigates the impact of the concentration of fibronectin in a solution used to functionalise polyacrylamide (PA) substrates on cell behaviour.
Methods: Mouse embryonic fibroblasts (MEF 3T3) were cultured on PA with elasticities of 20 and 40 kPa, functionalised with fibronectin in concentrations of 1, 5, 10, and 20 μg/ml.
Two-hour time-lapse brightfield microscopy was employed to acquire images of migrating cells. The resulting sets of images were used to quantify the cell-spread area and migration velocity. Based on these parameters and observed migration strategies, the cells were classified into distinct subpopulations.
Results: Four distinct cell subpopulations were identified: mesenchymal, amoeboid, slow amoeboid and polygonal/bigonal. Cells grown on softer substrates showed greater sensitivity to fibronectin concentration than those on stiffer substrates. The cell area was maximal and the velocity minimal for cells grown on PA substrates functionalised with 10-μg/ml fibronectin, making their parameters most similar to cells observed on glass. Cells underwent spontaneous transitions between subpopulations, although no direct transitions between mesenchymal and amoeboid or slow amoeboid phenotypes were observed.
Conclusions: Our findings highlight that cellular heterogeneity is not exclusive to cancerous cells and can also be observed in normal fibroblasts. Cells also can transit between them spontaneously. Moreover, cell behaviour is significantly influenced by the concentration of fibronectin in the functionalising solution, with a greater impact on softer substrates.
Objective: This study aimed to evaluate the impact of ionising radiation on exosome stability. We hypothesised that indirect effects of hydroxyl radicals would cause dose- and doserate-dependent structural modifications to extracellular vesicles (EVs).
Materials: Exosomes from PC3 prostate cancer cultures isolated via size exclusion chromatography (SEC) were irradiated using a Varian TrueBeam accelerator within a custom 3D-printed phantom equipped with EBT3 film dosimetry.
Methods: A two-stage Monte Carlo (MC) simulation was used to estimate radical G-values (molecules/100 eV) for 6 MV photon, 6 MeV electron, and 180 MeV proton beams using the Geant4-DNA toolkit. NTA quantification of CMDR-labelled vesicles was conducted following irradiation with the photon beam. Baseline EV presence before the photon irradiation was confirmed via AFM and cryo-EM.
Results: The MC simulations demonstrated that while the time-dependent curves followed characteristic evolutionary tracks, the final pool of hydroxyl (⋅OH) radicals at the end of the chemical stage did not differ significantly among the evaluated modalities, stabilising at ≈ 2.7 molecules/100 eV. The post-irradiation NTA (nanoparticle tracking analysis) revealed clear dose- and rate-dependent shifts manifested as distinct alterations in both the vesicle concentration and average size distributions. Exposure to ionising radiation resulted in a measurable increase in the average size of the exosomes. The newly proposed dose rate factor (DRF), defined as the ratio of relative change of the mean particle size for a given deposited dose delivered at various rates, maintained a stable value of 2.5 ± 0.3, closely
aligning with the physical ratio of instantaneous dose rates (≈ 2.33).
Discussion: Since the final G-values converged, the results demonstrate that while the cumulative abundance of radicals is determined by the total dose, the instantaneous dose rate dictates their local concentration, directly driving the kinetics of radical-mediated membrane modifications and cross-linking events.
Conclusions: This study demonstrates that while the total radiation dose governs the cumulative radical yield (G⋅OH ≈ 2.7 molecules/100 eV across all beams), the instantaneous dose rate dictates local radical concentrations and drives structural modifications of exosomes. This relationship can be quantified by the newly proposed DRF = 2.5 ± 0.3, which closely scales with the ratio of instantaneous dose rates (≈ 2.33). By establishing this direct link between Geant4-DNA
simulations and experimental NTA data, the findings provide a predictive, quantifiable framework for examining extracellular vesicle radiosensitivity and are thus directly relevant to the development of vesicle-based therapeutics and FLASH radiotherapy.
Objective: Quantitative assessment of muscle magnetic resonance imaging (MRI) is crucial for monitoring neuromuscular disorders (NMDs). This study introduces an automated radiomic phenotyping framework for fat replacement in the lower limbs, based on original features engineered across five main architectural domains: quantitative morphometry, spatial distribution, geometric shape, interactions between progressive fat replacement stages, and graph-based topology.
Material: Using 1184 T1-weighted (T1w) MRI scans, each corresponding to a unique patient from the CoMPaSS-NMD project, we mapped the complex three-dimensional (3D) architecture of heterogeneous intramuscular lipodegeneration to objective, morphologically interpretable biomarkers.
Methods: We introduced a graph-based skeletonisation of fat infiltrates to quantify muscle architectural changes, establishing a multidimensional extension of traditional, spatially agnostic volume metrics by mapping topological networks across the entire 3D muscle volume.Results: Non-parametric Kruskal–Wallis analysis confirmed the discriminative power of these novel descriptors across the genetic hierarchy. Notably, topological network metrics demonstrated large effect sizes (e.g., SF1_Skel_Nodes, ϵ2 = 0.2656), significantly outperforming classical global fat assessment (GF_Global_Pct, representing overall fat percentage), which yielded only a medium effect size (ϵ2 = 0.1172).
Discussion: These findings demonstrate that 3D topological profiling provides deeper structural insights than baseline volumetry.Post hoc pairwise evaluations and UMAP projections further indicated that these topological and 3D geometric invariants could capture disease-specific macroscopic infiltration patterns.
Conclusions: These results demonstrate that global architectural features represent a promising class of potential biomarkers for differential diagnosis. Future integration into clinical practice will require prospective evaluation on independent cohorts. The automated feature extraction pipeline developed in this study is integrated into and available within the MUSCAT (MUSCle fAt Topology) library.
From Macroscopic Dose to Radical Yield: A Multiscale Simulation and Experimental Study of Exosome Irradiation
- 1. Faculty of Physics, University of Warsaw, Warsaw, Poland
- 2. Department of Immunology, Medical University of Warsaw, Warsaw, Poland
- 3. National Medical Institute of the Ministry of the Interior and Administration, Warsaw, Poland
Abstract
Objective: This study aimed to evaluate the impact of ionising radiation on exosome stability. We hypothesised that indirect effects of hydroxyl radicals would cause dose- and doserate-dependent structural modifications to extracellular vesicles (EVs).
Materials: Exosomes from PC3 prostate cancer cultures isolated via size exclusion chromatography (SEC) were irradiated using a Varian TrueBeam accelerator within a custom 3D-printed phantom equipped with EBT3 film dosimetry.
Methods: A two-stage Monte Carlo (MC) simulation was used to estimate radical G-values (molecules/100 eV) for 6 MV photon, 6 MeV electron, and 180 MeV proton beams using the Geant4-DNA toolkit. NTA quantification of CMDR-labelled vesicles was conducted following irradiation with the photon beam. Baseline EV presence before the photon irradiation was confirmed via AFM and cryo-EM.
Results: The MC simulations demonstrated that while the time-dependent curves followed characteristic evolutionary tracks, the final pool of hydroxyl (⋅OH) radicals at the end of the chemical stage did not differ significantly among the evaluated modalities, stabilising at ≈ 2.7 molecules/100 eV. The post-irradiation NTA (nanoparticle tracking analysis) revealed clear dose- and rate-dependent shifts manifested as distinct alterations in both the vesicle concentration and average size distributions. Exposure to ionising radiation resulted in a measurable increase in the average size of the exosomes. The newly proposed dose rate factor (DRF), defined as the ratio of relative change of the mean particle size for a given deposited dose delivered at various rates, maintained a stable value of 2.5 ± 0.3, closely
aligning with the physical ratio of instantaneous dose rates (≈ 2.33).
Discussion: Since the final G-values converged, the results demonstrate that while the cumulative abundance of radicals is determined by the total dose, the instantaneous dose rate dictates their local concentration, directly driving the kinetics of radical-mediated membrane modifications and cross-linking events.
Conclusions: This study demonstrates that while the total radiation dose governs the cumulative radical yield (G⋅OH ≈ 2.7 molecules/100 eV across all beams), the instantaneous dose rate dictates local radical concentrations and drives structural modifications of exosomes. This relationship can be quantified by the newly proposed DRF = 2.5 ± 0.3, which closely scales with the ratio of instantaneous dose rates (≈ 2.33). By establishing this direct link between Geant4-DNA
simulations and experimental NTA data, the findings provide a predictive, quantifiable framework for examining extracellular vesicle radiosensitivity and are thus directly relevant to the development of vesicle-based therapeutics and FLASH radiotherapy.
LIST OF ABBREVIATIONS
AAA – analytical anisotropic algorithm
AFM – atomic force microscopy
CCB – Cyclotron Centre Bronowice
CDF – cumulative distribution function
CDFs – cumulative distribution functions
cryo-EM – cryogenic electron microscopy
DRF – dose rate factor
EVs – extracellular vesicles
FF – flattening filter
FFF – flattening filter-free
FUW – Faculty of Physics, University of Warsaw
GPR – gamma passing rates
MC – Monte Carlo
NTA – nanoparticle tracking analysis
PLA – polylactic acid
ROI – region of interest
SEC – size exclusion chromatography
SEM – standard error of the mean
BRIEF DESCRIPTION OF THE WORK
This study demonstrates how different radiation modalities induce radical-mediated structural modifications and concentration reductions in prostate cancer-derived exosomes. By linking multiscale Geant4-DNA modelling with experimental NTA (nanoparticle tracking analysis) data, the findings offer crucial insights into vesicle stability and dose-rate phenomena, providing a foundation for the development of future vesicle-based therapeutics and radiotherapy biomarkers.
INTRODUCTION
Exosomes are membrane-bound extracellular vesicles (EVs) secreted by virtually all cell types. EVs act as primary mediators of intercellular communication [1–4] by carrying ‘cargo’ such as proteins or nucleic acid fragments produced by the parent cell [5]. While EVs are essential for physiological homeostasis, they also play critical roles in various pathological processes, including the formation of premetastatic niches by tumour-derived vesicles [6, 7].
Since EVs are non-nucleated and lack the capacity to proliferate, their response to ionising radiation bypasses the classical pathway of reproductive cell death, being instead driven by direct and indirect structural alterations. Damage to the lipid membrane and subsequent exosome aggregation can be caused by interactions with radicals generated during water radiolysis, with the hydroxyl radical (⋅OH) playing a pivotal role due to its high reactivity [8]. However, the precise mechanisms underlying such interactions remain difficult to observe experimentally due to the nanometric scale of the vesicles. In this context, computational modelling has become an indispensable tool. Monte Carlo (MC) simulations enable bridging the gap between macroscopic dose delivery and microscopic molecular events, providing insight into the spatial distribution and density of ionising events that govern the production of radicals.
The stability of EVs is determined by environmental and storage conditions post-isolation. Factors such as storage temperature, duration before analysis, and repeated freeze-thaw cycles are well-established determinants of EV structural integrity, size distribution, and concentration [9]. Although the impac ts of these factors have been extensively documented, the effects of exposure to ionising radiation on EV stability and membrane properties remain relatively unexplored. Furthermore, while current research primarily focuses on exosomes derived from pre-irradiated cells, there is a notable lack of studies investigating the direct effect of radiation on isolated vesicles. To address this gap, our study introduces a novel approach by evaluating the direct irradiation of isolated EVs.
This study investigated whether exosome clustering could be detected via NTA following irradiation with photon beams, at doses of 2, 10, and 50 Gy delivered at different rates. A multiscale approach was employed to elucidate the underlying physical and chemical mechanisms. Geant4 and Geant4-DNA MC simulations were utilised to quantify the microscopic distributions of chemical radicals, offering a mechanistic explanation for the observed effects of different types of radiation.
The methodologies employed for exosome isolation and characterisation were in accordance with the MISEV2023 guidelines proposed by the International Society for Extracellular Vesicles [10]. The experimental workflow involved PC3 cell culture followed by exosome isolation via size exclusion chromatography (SEC). The presence, morphology, and structural integrity of the isolated vesicles were confirmed using atomic force microscopy (AFM) and cryogenic electron microscopy (cryo-EM). Finally, the samples were irradiated within a custom 3D-printed PLA phantom, and the results were validated through Gafchromic film dosimetry.
MATERIALS AND METHODS
Phantom preparation
The phantom was designed using the web tool Autodesk Tinkercad [11]. The design was imported into the FlashPrint 5 software to create the model and generate the output for the Flashforge Guider II S 3D printer. Printing was performed using transparent Spectrum PLA Premium filament, 1.75 mm, 1 kg (NATURAL, 5903175657091). The phantom was printed with 100% infill to eliminate internal air gaps.
The design of the phantom is shown in Fig. 1. The central opening was shaped to allow the walls of a 1.5 mL sample tube to fit snugly against the phantom. A rectangular pocket at the edge of the phantom was included to accommodate radiochromic films.
Fig. 1. The design of the phantom, with marked beam directions and the isocentre (black cross). All dimensions listed are in mm.
MC Simulations
MC simulations were performed in Geant4 software (ver. 11.3.1) [12–14]. The simulations were conducted in two distinct stages.
In the first stage, the entire phantom geometry was recreated in the simulation code. The simulation considered the phantom, the sample tube material, its contents (approximated using water), and the radiochromic film, divided into its ac tive and protective layers. The surrounding environment of the phantom was air. QGSP_BIC was used as the reference physics list for this stage of the calculations, with the G4EmLivermorePhysics constructor responsible for modelling electromagnetic processes for photons and electrons.
The electron density of the materials comprising the experimental setup was calculated based on the mass-fraction elemental composition and the mass density using the following formula:
where ρ is the mass density of the material; NA is the Avogadro constant; wi is the fraction by weight, Zi is the atomic number, and Ai is the atomic mass of the i-th element.
To simplify the dosimetric comparisons, we used the relative electron density (RED), defined as the ratio of the material’s ED to that of pure water. The composition and densities of the materials used in the simulation are listed in Tab. I.
Tab. I. Descriptions of the materials simulated in the first stage of the Geant4 calculations.
Particle transport was simulated using three source geometries corresponding to the experimental setups, with 4 × 107, 107 and 2 × 107 primary particle histories simulated for photons, electrons, and protons, respectively. To ensure internal consistency, all simulations for each configuration were performed in duplicate. For the photon and electron beams, the axis was oriented parallel to the film plane to obtain characteristic depth-dose profiles. In contrast, for the proton beam, the axis was oriented perpendicular to the film plane to capture the lateral dose profile, as irradiation parallel to the film would produce an almost uniform dose distribution within the phantom, corresponding to the entrance plateau region. Photons were emitted from a point source located 100 cm from the isocentre, which was positioned in the bottom part of the sample tube, as indicated by the cross in Fig. 1. In contrast, electrons were generated from a single uniform planar source with dimensions of 75 × 75 mm², positioned 5 cm from the phantom’s surface. The field size was limited compared to the experimental setup to reduce the calculation time. However, this margin allowed for the majority of electrons scattered into the phantom interior that would otherwise travel outside the direct field of view. Protons were simulated from the 30 × 30 mm² plane, positioned 2.5 mm from the phantom’s surface. Since the field was composed of a 13 × 13 spot map with a fluence shaped as a 2-dimensional Gaussian distribution (σ = 3.5 mm) and a spot spacing of 2.5 mm, the probability map was generated based on the superposition of 169 Gaussian functions distributed according to the spot map (Fig. 2.).
Fig. 2. Visualisation of the proton beam fluence consisting of a 13 × 13-spot grid, overlaid on the phantom wall. The cross marks the beam isocentre. The white dotted rectangle indicates the analysed film area (ROI), while the black dotted line indicates the edge of the radiation field.
The energy spectra for the simulated primaries for photon and electron beams were derived from published data [15, 16]. A cumulative distribution function (CDF), P(E), was constructed for each energy distribution. Primary particle energies were then sampled using the inverse transform method with a pseudorandom number generator p ∈ [0,1). To account for the discrete nature of the source data, linear interpolation was applied between adjacent points (Pi ; Ei) and (Pi+1 ; Ei+1) for Pi < p < Pi+1. For protons, the beam energy was sampled from a Gaussian distribution with μ = 180 MeV, σ = 0.5 MeV.
In this stage, the complete experimental setup was modelled using the Livermore physics list to track particle transport and extract the energy spectra of charged particles, both primary and secondary, entering the target volume, as well as the deposited dose. The target was defined as a cubic water voxel with a side length of 1 mm, located in the bottom part of a sample tube (Fig. 3.).
Fig. 3. Cross-section of the experimental setup. The red dot indicates the position of the target.
The energy distribution of the entering particles was used to construct a CDF for energy sampling in the second stage. The production threshold for secondary particles was set to 10 eV to ensure high precision in tracking low-energy interactions. This comprehensive energy spectrum served as the physical input for the subsequent microdosimetric phase.
In the second s tage, a dedicated simulation w as developed using the Geant4-DNA library [17–21]. The microscopic voxel identified as the target in the first stage was scaled up to define the entire simulation world, modelled as a cubic volume of water with a side length of 100 μm. Primary particles in this second phase were generated based on the energy spectrum obtained from the first stage, using the same CDF- -based approach. The generated particles were initiated at the centre of the voxel’s proximal face, directed perpendicularly toward the interior of the volume. For all configurations, five independent simulations of 100 primary particles each were conducted to ensure internal consistency.
Physical interactions were modelled using the G4EmDNAPhysics_ option4 constructor, while the chemical stage was simulated using G4EmDNAChemistry_option3. Due to the upper energy limits of the Geant4-DNA models, electrons with energies exceeding 1 MeV and protons with energies above 100 MeV were capped and simulated at these respective maxima for the allowable threshold values.
This high-resolution approach was employed to calculate the absorbed dose and the yield of radicals produced via water radiolysis and subsequent molecular interactions within the microscopic volume. This framework allowed for a realistic coupling between the energy deposition and the chemical yields. Based on the simulation results, the G-value, defined as a measure of the number of radiolytic species present per 100 eV of deposited energy, was determined for seven species (·H, ·OH, e–aq, OH–, H3O+, H2O2, and H2). Their dynamics were simulated between 10–3 and 103 ns (from ps to μs), corresponding to the chemical phase.
Validation and irradiation processes
The results of the first simulation stage were validated against the experimental measurements. The 3D-printed phantom described above containing a sample tube and radiochromic film was irradiated using 6 MV FF (with flattening filter) and 6 MV FFF (flattening filter-free) photon beams as well as a 6 MeV electron beam using the Varian TrueBeam linear accelerator (Varian Medical Systems, Palo Alto, CA, USA) at the National Medical Institute of the Ministry of the Interior and Adminis tration in Warsaw, Poland (PIM MSWiA), with a field size of 100 × 100 mm2. The proton beam was delivered using the Proteus-235 cyclotron (IBA, Louvain-la-Neuve, Belgium) at the Cyclotron Centre Bronowice in Cracow, Poland (CCB), following the field size and delivery technique described in the previous section. The beam energy at the phantom’s proximal surface was adjusted to 180 MeV.
The dose profiles obtained from EBT3 radiochromic films in the experimental setup were compared with the dose distributions simulated in Geant4. This comparison was performed to ensure the accuracy of the macroscopic energy deposition before proceeding to the microdosimetric stage.
In the experimental setup, the dose delivered to the target was 2 Gy, as calculated by the Varian Eclipse Treatment Planning System ver. 16 (TPS) used in clinical practice. The TPS calculations were verified using an IBA FC65-G farmer-type ionising chamber and an RW3 solid water slabs phantom. For the photon beams, the dose was delivered using parallel-opposed fields, with the total number of monitor units (MUs) divided equally between the two opposing sides to ensure a balanced dose distribution. The dose rates for the FF beam were 600 MU/min (≈ 7.5 Gy/min at isocentre) and 1400 MU/min (≈ 17.5 Gy/min at the isocentre) for the FFF beam. A single field was used for the remaining beam types. The detailed irradiation setup with beam directions is presented in Fig. 1.
Due to the high computational demand of the MC simulations, the number of primary particle histories in Geant4 was limited, resulting in lower values of doses deposited in the target (order of 10-4 Gy, with relative standard uncertainty of ≈ 1%). To allow for direct comparison between the experimental and numerical results, the simulated dose distributions were normalised such that the mean absorbed dose within the target volume corresponded to the nominal value of 2 Gy.
Calibration of the radiochromic film was performed using a clinical 6 MV FF photon beam. Standard procedures for film calibration, processing, and dose readout were followed, as described in our previous work [22].
To evaluate the dose distribution, dose profiles were extracted from the radiochromic films aligned along the beam y-axis (Fig. 3.). The analysis was performed using a rectangular region of interest (ROI) with a constant width of 10 mm, marked as analysed film area. To enhance the signal-to-noise ratio, pixel values were averaged laterally (across the 10 mm width) for each depth increment. This procedure yielded a mean absorbed dose and its corresponding standard deviation as a function of the distance from the phantom axis, providing a robust dataset for comparison with the MC-simulated profiles.
The simulated dose distributions were analysed using an identical numerical approach. In the Geant4 environment, the radiochromic film volume was discretised into a two-dimensional matrix, allowing for pixel-by-pixel dose recording. The dose values were averaged across a 10 mm lateral span for each simulated depth, mirroring the experimental ROI processing. This consistency in data reduction ensured a direct and unbiased comparison between the measured and calculated dose profiles. In addition, the data collected within the range of the film width were resampled to a uniform step of 0.1 mm using linear interpolation to ensur e consistency across datasets with varying resolutions.
In the Geant4 simulations, the scoring matrix was defined with a pixel size of 0.5 mm, whereas for the experimental validation, the EBT3 films were scanned at a resolution of 150 DPI (corresponding to a pixel size of approximately 0.17 mm).
Cell culture and exosome isolation
Human prostate cancer PC3 cells (CRL-1435) were cultured at the Radiobiological Laboratory at the Faculty of Physics, University of Warsaw (FUW) according to the established laboratory protocol described in a previous study [23]. All experiments were conducted under identical culture conditions as reported in the referenced article. The cells were maintained in 75 cm² tissue culture flasks (VWR, 734-2313) in DMEM/F-12 medium (VWR, 392-0411) enriched with 10% FBS and 2% penicillin–streptomycin (Gibco, 15140-122).
Standard FBS (Biowest, S1810-500) was utilised up to the fourth passage, after which the culture medium was transitioned to one supplemented with exosome-depleted FBS (Gibco, A27208-01) to prevent contamination of isolated samples with vesicles originating from the serum. Conditioned medium for the isolation of exosomes was collected starting from the second passage following this transition.
Exosomes were isolated according to a protocol optimised for vesicles collected from in vitro cell cultures [24]. Briefly, the conditioned medium was sequentially filtered using Millex filters (MilliporeSigma, SLGP033RS), concentrated with Vivaspin centrifugal spin columns (Sartorius, VS2041) with 100 kDa MWCO, and further purified via SEC using Sepharose resin (Cytiva, GE17-0140-01). The exosomes were eluted with Dulbecco’s phosphate-buffered saline (DPBS) (Gibco DPBS, 14190-136). To ensure a uniform exosome concentration across all conditions, all collected fractions were pooled into a single stock solution and subsequently aliquoted into 1 mL por tions using 1.5 mL Eppendorf tubes, yielding a total of 27 identical samples. Three non-irradiated tubes were dedicated to the baseline control characterisation methods described in subsequent subsections, with each assay (AFM, cryo-EM, and non-irradiated NTA baseline) utilising an independent tube. Seven tubes were used for the primary irradiation experiments and post-exposure NTA analysis reported in this study, with three independent measurement replicates performed for the material from each tube. The remaining 17 tubes were allocated to additional experimental conditions not included in the scope of this work. All samples were stored at 4°C and handled within 48 h post-preparation.
Cryogenic electron microscopy
After isolation, the morphology and s tructural integrity of the exosomes were verified using a Thermo Scientific 200kV Glacios – Cryogenic Electron Microscope (cryo-EM) at the Cryomicroscopy and Electron Diffraction Core Facility, Centre of New Technologies, University of Warsaw (CeNT UW). Unlike conventional TEM, cryo-EM allows for the observation of biological specimens in their native, hydrated state. The sample was concentrated using Amicon® Ultra-0.5 Centrifugal Filter Units with a 100 kDa MWCO (Merck Millipore, Cat. No. UFC510096) at a ratio of 1:50 in filtered DPBS. A 3 μL aliquot of the concentrated sample was applied onto a copper Quantifoil R2/1 grid (Cu, 2 nm carbon film) that had been previously glow-discharged. Vitrification was performed by plunge-freezing using a Thermo Fisher Scientific Vitrobot Mark IV under the following controlled conditions: a preparation temperature of 19°C, 100% relative humidity, a blot time of 4 s, a blot force setting of 4, and a wait time of 10 s. Imaging was conducted using a Thermo Fisher Scientific Glacios TEM operated at an accelerating voltage of 200 kV and equipped with a Falcon 3EC direct electron detector (4096 × 4096 pixels). Micrographs were acquired at a magnification of 92000×, corresponding to a calibrated pixel size of 1.587 Å at the specimen level. The defocus value was set to –3 μm, with a total electron dose of approximately 50 e−/Å2. Under these conditions, exosomes should appear as spherical, membrane-bound structures consistent with the characteristic size and morphology, with a clearly visible lipid bilayer approximately 5 nm in thickness [25].
Atomic force microscopy
EVs were imaged in DPBS on APTES-modified mica using a Dimension Icon AFM (Bruker, Santa Barbara, CA, USA) operating in PeakForce QNM mode. Topographical maps were acquired over a 3 μm × 3 μm area (256 × 256 pixels, ~12 nm resolution) at a 0.98 Hz line rate and 2 kHz modulation frequency, using a soft biological cantilever (PEAKFORCE-HIRS-F-A, Bruker) calibrated to a 43.5 kHz resonance frequency and 35.1 nm/ V deflection sensitivity. The images obtained were reviewed and analysed using Gwyddion 2.69 software [26].
Nanoparticle tracking analysis
The concentration and size distribution of the isolated vesicles were characterised using NTA with a ZetaView (Particle Metrix) instrument. To achieve the optimal particle density for tracking, the isolates were concentrated using Amicon® Ultra-2.0 Centrifugal Filter Units with a 100 kDa MWCO (Merck Millipore, Cat. No. UFC210024) and subsequently diluted at a ratio of 1:300 in filtered DPBS. The initial measurements were conducted in scatter mode (λlaser = 488 nm) to establish a baseline size distribution histogram for all detected particles. The samples were incubated with the lipophilic fluorescent dye CellMask™ Deep Red (CMDR, Thermo Fisher Scientific) to specifically identify and quantify membrane-bound vesicles. The labelled samples were then re-analysed using both scatter and fluorescence (stain, λlaser = 640 nm, λfilter = 660 nm) modes. This dual-mode approach allowed for the discrimination of biological exosomes from non-membranous background noise, providing a verified particle concentration for the subsequent irradiation experiments. The post-irradiation NTA analysis focused on comparing particle concentrations and mean sizes across seven experimental groups: an unirradiated control and samples irradiated with doses of 2, 10, and 50 Gy using both FF and FFF photon beams. To ensure statistical reliability, three independent replicates were recorded for each condition.
To quantify the biological fraction and structural changes induced by radiation, the following parameters were derived for each sample:
- Concentration (C): The particle concentration was assessed in both scatter and stain modes, with three measurements for each sample.
- Fluorescence Ratio (R): Calculated as the ratio of the mean particle concentration (C) in stain mode to the concentration in scatter mode (Cstain/Cscatter), representing the fraction of membrane-positive vesicles.
- Size Differential (ΔS): Calculated as the difference between the mean particle size (S) measured in stain mode and scatter mode (Sstain – Sscatter).
- Dose Rate Factor (DRF): DRF evaluates the impact of the dose rate on the size distribution of irradiated exosomes and was defined according to the following formula:
where D is the absorbed dose.
This comparative framework was designed to detect potential exosome aggregation and alterations in membrane integrity induced by varying dose rates and total absorbed doses.
Statistical analysis
In the first stage of the Geant4 simulations , the uncer tainty of the absorbed dose for each individual pixel in the analysed film area was estimated using the standard error of the mean (SEM) and was maintained below 3%. To derive the final dose profile, the mean absorbed dose and its associated uncertainty at each depth were calculated using a weighted average method, where the uncertainties in individual pixels served as the weighting factors.
In the case of radiochromic film dosimetry, the uncertainty of each individual pixel was derived from the confidence intervals of the fitted calibration function, obtaining the lowest uncertainty values below 2.5% for red channel analysis.
The TPS dose calculations were performed using the analytical anisotropic algorithm (AAA). Based on the known precision of this algorithm in homogeneous and near-homogeneous media, a 1% uncertainty was assigned to each calculated pixel.
The same weighted average method was subsequently applied to the datasets to determine the mean dose and its associated uncertainty at each depth, thereby ensuring a consistent comparison with the experimental and MC results.
A quantitative comparison between the measured and simulated depth-dose profiles was performed using the gamma index analysis (1 mm / 1% and 2 mm / 2%). A gamma pass r ate was then calculated to evaluate the degree of agreement between the experimental film data and the MC simulation results.
Two-sample Kolmogorov–Smirnov and Wilcoxon signed-rank tests were employed for comparative analysis of two datasets, whereas the Friedman test was utilised for evaluations involving larger sets of multiple related groups.
The relative uncertainty of particle concentration C in the NTA analysis was estimated as the inverse of the square root of the number of registered tracks, following Poisson statistics. The final concentration value and its associated uncertainty were determined by calculating the weighted average of three independent sample measurements, with the individual measurement uncertainties used as the weighting factors. The uncertainties for the parameters R, ΔS and DRF were subsequently determined using the law of propagation of uncertainty, accounting for the errors associated with the primary measured variables.
RESULTS
The NTA of pre-irradiation exosome samples produced reproducible concentration profiles among the evaluated replicates (Fig. 4B.). The particle size distributions consistently exhibited a single dominant peak, with the maximum concentration (mode) located at approximately 100 nm. Statistical evaluation of the three concentration histograms using the Friedman test indicated no significant differences in the size distribution curves (χ2F(2) = 1.459, pF = 0.482, N = 200 data points per v ector). A representative video frame captured from the NTA camera illustrated the high particle density and stable scatter signals recorded during the measurements (Fig. 4A.). Cryo-EM imaging (Fig. 4B.) revealed intact exosomes exhibiting the typical dimensions and a characteristic spherical morphology, with a clearly defined lipid bilayer. Furthermore, AFM topographical images (Fig. 4C.) displayed a large population of isolated objects distributed across the substrate exhibiting typical exosome dimensions.
Fig. 4. Pre-irradiation verification of exosome samples. (A) Left: Histograms of the concentrations for three measured samples. Right: Representative video frame capture from the NTA camera during measurement; (B) Representative cryo-EM images capturing individual vesicular structures within the characteristic exosome size range; (C) Representative AFM topographic images displaying a high density of structures within the typical exosome size range.
The dose profiles obtained from EBT3 measurements and the corresponding Geant4 MC simulations were highly consistent across all examined radiation types (Fig. 5.). The experimental and simulated curves overlapped within their respective uncertainty bands for photons (Fig. 5A.), electrons (Fig. 5B.), and protons (Fig. 5C.). A gamma index analysis was performed to quantitatively evaluate the agreement between the measured and calculated distributions; the resulting gamma passing rates (GPR) are summarised in Tab. II.
Fig. 5. Comparison of dose profiles obtained from Geant4 MC simulations and Gafchromic EBT3 film measurements for different radiation types. The panels represent profiles for photons (A), electrons (B), and protons (C) as a function of distance from the phantom axis. The dashed lines indicate the mean values for EBT3 films (blue) and Geant4 simulations (red). Shaded areas correspond to the respective experimental and simulation uncertainty bands.
Tab. II. Summary of the GPR for investigated beams.
The energy spectra of primary particles and the corresponding spectra of charged particles entering the target are presented alongside their CDFs in Fig. 6. The distributions are reported for 6 MV FF photons (Fig. 6A.), 6 MV FFF photons (Fig. 6B.), 6 MeV electrons (Fig. 6C.), and 180 MeV protons (Fig. 6D.). For the spectra of target-entering particles from the primary photon beams, a two-sample Kolmogorov–Smirnov test (D = 0.065, pKS = 0.158, N = 600) and a Wilcoxon test (W = 56479, pW = 0.627, N = 600) indicated no significant difference between the distributions. In contrast, for the remaining radiation modalities, there was significant variation in shape and peak position between the primary particle spectra and the spectra of secondary charged particles interacting with the target. To clarify the overlapping features and detailed variation within the proton distributions, a high-resolution close-up of two specifically selected energy regions from the 180 MeV proton dataset (Fig. 6D.) is shown in Fig. 6E. The low-energy region highlights a population of secondary electrons extending up to a maximum cutoff energy of approximately 0.43 MeV. Meanwhile, the high-energy region displays a symmetric proton peak characteristic of a Gaussian distribution centred at approximately 170 MeV, with an FWHM of approximately 1.3 MeV.
Fig. 6. Energy spectra of the primary particles presented alongside energy spectra of charged particles entering the target, including their cumulative distribution functions (CDFs). (A) 6 MV FF photons; (B) 6 MV FFF photons; (C) 6 MeV electrons, and (D) 180 MeV protons; (E) provides a close-up of two selected regions from panel (D).
The time-dependent evolution of G-values for the primary radiolytic species (·H radicals, ·OH radicals, and hydrated electrons, e–aq ) from 10−3 ns to 103 ns is presented in Fig. 7. There were distinct temporal dynamics and differences in magnitude across the specific chemical species for all three radiation modalities (photon, electron, and proton beams). The sums of positive and negative charges remained identical within the simulated volume to an accuracy of ≈ 1% throughout the chemical stage. Notably, the photon and electron beams were highly similar for each investigated species. The lack of statistically significant differences between their temporal trajectories was confirmed by a KS test for ⋅H (S = 0.2353, p = 0.119), ⋅OH (S = 0.0059, p ≈ 1), and e–aq (S = 0.0059, p ≈ 1) with N = 51. F or ·H radicals (Fig. 7A.), the proton beam consistently yielded a higher G-value (≈ 0.9 to 0.83) compared to the photon and electron beams, which exhibited overlapping trajectories at a lower level (≈ 0.7 to 0.62). All three beams showed a slight initial decrease in ·H yield up to ≈ 1 ns, followed by a minor recovery trend toward 103 ns.
Fig. 7. The temporal evolution of the G-values for ·H, ·OH and e– aq radicals for the investigated radiation types. Note that the scale of the left subfigure (A) is different from the others (B, C).
In the case of ·OH radicals (Fig. 7B.), the initial G-values at 10−3 ns ranged from ≈ 5.1 (photons/electrons) to 5.4 (protons). Beyond 10−2 ns, there was a sharp, continuous decline in yield across all radiation types, with the curves progressively converging to a nearly identical value of approximately 2.7 at 103 ns.
Similarly, the hydrated electron (e-aq ) yields (Fig. 7C.) started at around 4.1 to 4.2 at 10−3 ns and remained stable until approximately 10−2 ns. This was followed by a significant decay phase. In this region, a distinct divergence occurs where the proton beam yields the lowest e–aq values (≈ 2.5 at 103 ns), while the photon and electron beam curves remained tightly coupled, reaching a higher final yield of ≈ 2.7 at 103 ns.
The impact of radiation dose and dose rate on the physical properties of the isolated exosomes is summarised in Fig. 8. Panel A tracks the alterations in particle concentration (C), fluorescence ratio (R), and size differential (ΔS) across different dose levels (2, 10, and 50 Gy) for both standard FF and FFF irradiation modes, compared against the non-irradiated control (0 Gy).
The total particle concentration measured in scatter mode remained relatively stable across all groups, maintaining an order of magnitude of approximately 1010 mL−1. Conversely, the concentration detected in stain mode exhibited a dose-dependent decrease, declining under FFF conditions at higher doses down to the order of 109 mL−1 (Fig. 8A., top). Consequently, the fluorescence ratio R, which denotes the fraction of membrane-positive vesicles, demonstrated a continuous decline from its baseline value of ≈ 63% at 0 Gy. This reduction was more pronounced for the FFF mode than the FF mode at equivalent doses, reaching its lowest values (≈ 7%) at 50 Gy (FFF) (Fig. 8A., middle). There was a corresponding upward trend for the size differential ΔS (Fig. 8A., bottom). The difference in mean sizes obtained using the two measurement methods (fluorescence and scatter) increased systematically from an initial near-zero baseline of ΔS(0) ≈ (8 ± 8) nm at 0 Gy (indicated by the horizontal dashed line in the figure). Across all evaluated doses, the FFF mode induced a larger shift in ΔS compared to the FF mode, increasing to a maximum shift of (62 ± 9) nm at 50 Gy (FFF). DRF was computed to quantitatively evaluate the dose-rate relationship (Fig. 8B.). The calculated DRF values exhibited a stable, flat trajectory across the entire dose range, yielding values of 2.7 ± 0.7, 2.4 ± 0.5, and 2.6 ± 0.4 at 2 Gy, 10 Gy and 50 Gy, respectively. This consistency indicates that the structural modification effect induced by the high-dose- -rate FFF mode remained scaled by a constant factor relative to the standard FF mode, independent of the total absorbed dose, with a mean DRF value of 2.5 ± 0.3.
Fig. 8. Results of the NTA measurements of irradiated samples. (A) Values of the concentration (C), fluorescence ratio (R), and size differential (ΔS) parameters: red – scatter mode, blue – fluorescence mode. The x-axis represents the absorbed dose in Gy. The beam type is indicated in the brackets; (B) Values of the DRF parameters (see Formula 2) calculated for each dose. The orange bar represents the weighted mean of all three DRF values and its uncertainty.
DISCUSSION
The agreement between the experimental Gafchromic EBT3 dose profiles and the Geant4 MC simulations, as demonstrated by the high GPR value across all radiation modalities, confirmed the accuracy of the established computational framework. Achieving a GPR exceeding 96% even under the most stringent clinical evaluation criteria (1 mm / 1%) provides a rigorous foundation for evaluating the microscopic energy deposition tracks and primary particle spectra interacting with the exosomes. The spectral analysis (Fig. 6.) highlights the fundamental phy sical differences between the primary beams and the actual radiation fields entering the target. The altered shapes of the particle spectra reaching the targets are in agreement with theoretical expectations, directly reflecting the characteristic energy-loss mechanism inherent to the respective primary particle type. A key methodological finding emerged from the statistical comparison of the secondary electron spectra entering the target from the two different photon modes. The two-sample Kolmogorov-Smirnov and Wilcoxon tests confirmed that there was no statistically significant difference between the target-level electron distributions generated by the FF and FFF beams. This apparent equivalence justified the choice to streamline the subsequent analyses by focusing exclusively on a single representative photon-induced electron configuration. For the 180 MeV proton beam, the detailed high-resolution spectrum shown in Fig. 6E. displayed a characteristic symmetric Gaussian peak shifted down to approximately 170 MeV, resulting from the expected partial energy loss in the pre-target layers. The low-energy region identified in the figure indicated a distinct population of secondary electrons extending to a maximum cutoff of ≈ 0.43 MeV. This cutoff is highly representative of δ-electrons generated via knock-on collisions, moving at the maximum kinematically permissible energy transfer from a primary 170–180 MeV proton beam to atomic electrons. This electron field is key to creating a high-LET-like track structure and increasing radiolytic events.
The time-dependent evolution of radiolytic yields (G-values) from 10−3 ns to 103 ns exhibited tight error bands (Fig. 7.), highlighting the high statistical reliability achieved through the optimisation of the simulated primary particle histories. The total charge balance demonstrated that a negligible number of charged species left the simulation boundaries, thereby confirming that the defined geometry was sufficiently large to contain the evolution of tracks without boundary artefacts. A primary finding from the temporal trajectories is the statistical equivalence between the photon and electron beams across all evaluated species, as confirmed by the two-sample KS test. On a nanometer scale, both modalities deposit energy via sparsely ionising patterns (low-LET), thereby generating identical secondary electron fields that drive the chemical stage of radiolysis.
In contrast, the 180 MeV proton beam introduced distinct spatial features that significantly altered the chemical kinetics. There was a highly specific profile for the hydrogen radical yield (Fig. 7., top), which initially decreased up to 1 ns before recovering and stabilising. The initial drop in ⋅H yield was driven by intense radical intra-track recombination within the highly concentrated ionisation spurs, where ⋅H radicals either undergo homorecombination to produce molecular hydrogen (H2) or react with hydroxyl radicals to reform water molecules (H2O) . Beyond 1 ns, however, a secondary production channel becomes dominant as hydrated electrons react with hydronium ions (e–aq+H3O+→⋅H + H2O). In this later time domain, the rate of secondary ⋅H generation outpaces its consumption, causing the characteristic upward inflection and subsequent stabilisation of the curve.
This fast-kinetic interplay is mirrored by the decay profiles of both ⋅OH and e– aq (Fig. 7B., C.), which lack any secondary production mechanisms after the physical phase. The monotonic decline of the hydrated electron G-value is a direct consequence of its consumption in the aforementioned reaction with H3O+. This process is most pronounced for the proton beam, where the high localised density of reactants within the core of the track significantly accelerates e– aq consumption, resulting in a deeper drop in the proton curve compared to the other investigated beams.
Complementarily, the higher initial yield of ⋅H radicals for protons – which closely parallels a corresponding excess in initial ⋅OH yields – provides direct evidence of a higher rate of primary water molecule dissociation (H2O∗) during the physical stage. Despite this initial excess of ⋅OH radicals in the proton tracks, the curves for all three beams converged to an identical baseline (≈ 2.7) by 103 ns. The dense track structure of the proton beam places neighbouring ⋅OH radicals in close proximity, increasing the probability of their homorecombination into hydrogen peroxide (H2O2). This enhanced self-consumption effectively neutralises the initial proton-induced excess, ultimately equalising the pool of free hydroxyl radicals available at the end of the chemical stage, regardless of the primary radiation modality.
The macroscopic manifestations of these radiolytic and physical interactions are clearly visible in the dose-dependent structural degradation of the isolated exosomes (Fig. 8.). The pre-irradiation baseline characterisation (Fig. 4.) via NTA, cryo-EM, and AFM established a highly reproducible, intact population of spherical vesicles centred at ≈ 100 nm with well-defined lipid bilayers. Following irradiation, the dose-dependent decrease in the fluorescence ratio (R) in stain mode (declining from ≈ 63% to ≈ 7%), coupled with the preservation of total particle concentrations in scatter mode (≈ 1010 mL−1), indicates that the radiation did not completely disintegrate the physical vesicular envelope into scattered fragments. Instead, it systematically compromised membrane integrity or surface-staining epitope configurations, rendering the vesicles “membrane-negative”.
A plausible biochemical driving force contributing to these structural alterations is lipid peroxidation, initiated by radiolytically generated hydroxyl radicals (⋅OH) extracting hydrogen atoms from polyunsaturated fatty acid chains (LH+⋅OH→⋅L+H2O). In the presence of dissolved oxygen, this propagates a radical cascade forming peroxyl radicals (LOO⋅) and lipid hydroperoxides (LOOH), an alteration that disrupts fatty acid packing, decreases membrane fluidity, and alters the overall curvature. This free-radical-induced destabilisation leads to membrane rearrangement, vesicle-vesicle fusion, or swelling due to altered osmotic permeability, providing a plausible mechanistic explanation for the observed increase in mean exosome size.
This structural disruption is mirrored by the significant broadening of the size differential ΔS, which increased from a baseline value of ≈ 8.3 nm to a maximum of 62.3 nm under 50 Gy FFF conditions. While the increase in ΔS points toward radiation-induced aggregation, swelling, or structural deformation heavily mediated by highly reactive ⋅OH and e– aq radicals, several alternative explanations should also be considered. Specifically, the observed drop in the fluorescence-positive fraction alongside stable scatter counts could stem from radiation-induced membrane damage that impairs the binding affinity or retention of the CMDR dye.
Beyond structural alterations, these changes in membrane structure may have significant functional consequences for EV-mediated intercellular communication. On the one hand, vesicle clustering could concentrate the encapsulated cargo, potentially amplifying the local signal delivered to recipient cells. On the other hand, severe aggregation or structural deformation might hinder cellular uptake and recognition pathways, ultimately impairing the efficiency of EV-based information transfer.
The results indicate a prominent dose-rate effect. At each equivalent dose level (2, 10, and 50 Gy), the FFF mode characterised by an ≈ 2.33 higher instantaneous dose rate induced a stronger reduction in R and a larger shift in ΔS than the standard FF mode. When quantified via the newly proposed DRF, this phenomenon demonstrated a stable, flat trajectory across the entire dose range, maintaining a tight mean value of 2.53 ± 0.29. Remarkably, within the experimental uncertainties, this DRF value was in agreement with the physical ratio of the instantaneous dose rates between the two modalities, supporting the link between macro-scale structural damage and micro-scale beam delivery kinetics.
The independence of the DRF from the total absorbed dose implies that the higher dose rate did not alter the fundamental chemical nature of the individual radical-biomolecule reactions. Instead, the high dose scaled the overall structural damage by a constant factor. In FFF mode, the massive temporal compression of primary particle delivery creates a high spatio-temporal overlap of radical tracks. This overlap leads to a local concentration of oxidative species that exceeds the immediate radical-scavenging capacity of the surrounding solution, thereby accelerating collective oxidative stress and cross-linking events on the exosome membranes. The constancy of the DRF validates it as a predictable, scalable parameter that could be highly useful for optimising dose-rate strategies in future therapeutic or bio-engineering applications involving EVs.
The temporal trends and absolute G-values obtained herein are in agreement with benchmarks established in the literature. The characteristic shapes of the time-dependent curves and the evolution profiles for each of the investigated radiolytic species closely mirror those reported by Baba et al. (2021) in their quantitative Geant4-DNA analysis of water radiolysis [27]. Comparing specific magnitudes with other studies, Kusumoto et al. (2021) reported an initial yield of ≈ 5 molecules/100 eV for carbon beams at 10−3 ns post-irradiation, while Jay-Gerin et al. (2000) observed 4.6 molecules/ 100 eV at 10−1 ns for a 300 MeV proton beam [28, 29]. This aligns with the early-phase values (≈ 5.1 to 5.4 molecules/100 eV), capturing the highly concentrated state of primary products before track expansion dominates. In contrast, lower values such as 2.67 molecules/100 eV have been obtained by Meesungnoen et al. (2001) at longer time scales (e.g., 103 ns) for 137Cs gamma rays [30]. This long-term drop mirrors the observed convergence toward a baseline of ≈ 2.7 molecules/100 eV at 103 ns. The consistency between the simulated curves and the results of these independent models validates the absolute scale of the chemical stage kinetics, confirming that the micro-scale radical yields driving the subsequent exosome degradation are robustly quantified.
Several limitations of the present study should be acknowledged. First, the experimental findings are based on EVs isolated from a single cell line using one purification workflow; thus, the generalisability of these responses across different cell sources and isolation methods remains to be established. Second, while the baseline EV characterisation was confirmed via AFM, cryo-EM, and NTA, post-irradiation evaluation was limited to NTA quantification, and future work would benefit from complementary structural and molecular analyses. Third, although photon, electron, and proton irradiation setups were experimentally established and dosimetrically validated, direct experimental evaluation of the dynamics of the EV concentration and size distribution was restricted to the photon FF and FFF beams. Verifying the influence of the remaining types of radiation is the next phase of our research pipeline. The simulated differences in radiolysis yields among photon, electron, and proton interactions were not experimentally verified at the EV level; this could be a valuable extension of our research. Finally, the relationship between the DRF and the dose-rate ratio currently represents an empirical observation that requires rigorous independent validation across broader experimental conditions.
CONCLUSIONS
This study provides a comprehensive, multiscale examination of radiation-induced modifications of EVs, evaluating the interplay between micro-scale radiolysis and macro-scale structural changes.
NTA made it possible to observe noticeable shifts in the characteristics of irradiated exosomes. Interestingly, while the simulation results indicated that the final pool of hydroxyl (⋅OH) radicals at the end of the chemical stage did not differ significantly among the radiation modalities, the experiments indicated clear dose- and doserate- based effects manifested as distinct impacts of these parameters on both the concentration and average size distributions of the CMDR-labelled vesicles. The newly proposed DRF, defined as the ratio of the relative change in mean particle size observed for a given deposited dose delivered at different rates, maintained a stable value of 2.5 ± 0.3, in close agreement with the physical ratio of the instantaneous dose rates (≈ 2.33), suggesting a predictable scaling of these structural modifications. The developed simulation framework is fully prepared to integrate new physical and chemical constructor lists currently being developed within the Geant4-DNA toolkit; this will allow for even higher modelling precision in the future.
Ultimately, these findings contribute to a better quantitative understanding of dose-rate phenomena in biomimetic systems. Future work should focus on exploring these mechanisms under ultra-high-dose-rate (FLASH) conditions, where the biological and structural responses of EVs could reveal interesting kinetics relevant to modern radiotherapy.
CONFLICT OF INTEREST
The authors have no potential conflicts of interest to declare.
DATA AVAILABILITY
The raw and processed data, along with the simulation protocols generated during this study, are currently not publicly available, as they form part of an ongoing PhD dissertation. However, they can be made available by the corresponding author upon reasonable request.
FUNDING
All funding sources for the research were from the budget of the Faculty of Physics, University of Warsaw.